5-6 DELTA/DELTA XL/KAPPA VF8
5 A
LARMS
Code: Press once to silence alarm tone (in network mode) or reduce volume to
10 % (in standalone mode) and activate and display an event timer. Press
again to deactivate all active Code functions. Press a third time to deactivate the
event timer (see page 2-17 for more information).
Alarm Control
Many alarm functions, including alarm suspension, validation, and the display of
alarm limits, can be configured only on the Alarm Control menu, which in turn is
accessible only via the password-protected Unit Manager menu. For a description of
available functions on the Alarm Control menu, see page 2-13.
Alarm Setup (Alarm Limits Table)
The alarm limits table allows you to modify the alarm limits of multiple parameters in
a single location.
The Alarm Limits table displays values only if the associated parameter has been
prioritized (see page 2-5) or the associated monitoring device (for example, the NBP
cuff or the etCO
2
/respiratory mechanics pod) is connected.
Press the Alarm Limits fixed key. The Alarm Limits table appears.