14-20 DELTA/DELTA XL/KAPPA VF8
14 I
NVASIVE BLOOD PRESSURE
Status Messages
Message Possible Cause Suggested Action
<xx> S <#>
<xx> D <#>
<xx> M <#>
The pressure value is outside the
alarm limits.
Check the patient and treat, if necessary.
Access the Alarm Limits menu and change
the alarm limits.
Check the equipment and replace it, if
necessary.
<xx> Out of
Range (High)
<xx> Out of
Range (Low)
The pressure signal is out of the
measurement range.
Check the patient and treat, if necessary.
Access the Pressure Labels menu and
assign the correct label.
Check the equipment and replace it, if
necessary.
<xx> Please
Check Zero
The IBP zero value stored in the
monitor may not correspond to
the peripheral device.
Zero the transducer.
<xx> Static
Pressure
Static pressure was detected on a
pulsatile signal, owing to one of
the following conditions:
A physiological condition, for
example, asystole.
The transducer was turned off to
the patient.
A catheter tip lodged against a
vessel wall.
A clot on the catheter tip.
Check the patient and treat,if necessary.
Open the system to the patient by turning the
stopcock.
Follow hospital procedures for dislodging
catheters.
Follow hospital procedures for clotted
catheters.
<xx>
Unplugged
The pressure transducer for the
specified parameter is either
unplugged or defective.
Active pressure: reconnect or replace the
cable.
Inactive pressure: turn off alarms.
<xx> Zero
Required
The pressure transducer for the
specified parameter requires
zeroing.
Zero the transducer.
<xx> Zero
Accepted
The zeroing of the transducer was
successful
No action is required.
<xx> Did Not
Zero
The zeroing of the transducer
failed because of one of the
following conditions:
Excessive signal noise
A non-static waveform
Keep all tubing motionless, then rezero.
Change the transducer.
Check the stopcock, then rezero.
<xx> Did Not
Zero - Offset
Error
The zeroing of the transducer
failed because static pressure
was too high or too low.
Rezero the transducer.
Loosen and retighten the transducer dome,
then rezero the transducer.
Replace the transducer.
NOTE: <xx> represents the IBP parameter label associated with the displayed message.

Table of Contents

Other manuals for Dräger Infinity Kappa

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Dräger Infinity Kappa and is the answer not in the manual?

Dräger Infinity Kappa Specifications

General IconGeneral
BrandDräger
ModelInfinity Kappa
CategoryMedical Equipment
LanguageEnglish

Related product manuals