5-8 DELTA/DELTA XL/KAPPA VF8
5 A
LARMS
Upper and Lower Alarm Limits
Alarm limits should be set according to your patient’s prevailing condition within the
predefined ranges of the monitor listed in the following table.
Parameter Predefined Alarm Range
Default
State
Default Alarm
Setting / Alarm Priority
ARR See “Arrhythmia Setup Table” on page 9-6.
ART S/M/D -5 to 300 mmHg (-0.7 to 40.0 kPa) Off S: (Adult)
Low 90 mmHg (12.0 kPa)
High 160 mmHg (21.3 kPa)
M: (Adult)
Low 60 mmHg (8.0 kPa)
High 125 mmHg (16.7 kPa)
D: (Adult)
Low 50 mmHg (6.7 kPa)
High 110 mmHg (14.7 kPa)
S: (Pediatric)
Low 50 mmHg (6.7 kPa)
High 120 mmHg (16.0 kPa)
M: (Pediatric)
Low 50 mmHg (6.7 kPa)
High 80 mmHg (10.7 kPa)
D: (Pediatric)
Low 30 mmHg (4.0 kPa)
High 80 mmHg (10.7 kPa)
S: (Neonatal)
Low 50 mmHg (6.7 kPa)
High 120 mmHg (16.0 kPa)
M: (Neonatal)
Low 40 mmHg (5.3 kPa)
High 85 mmHg (11.3 kPa)
D: (Neonatal)
Low 35 mmHg (4.7 kPa)
High 80 mmHg (10.7 kPa)
BIS (From
BISx pod only.
You cannot
control alarms
from Aspect
A2000)
10 to 100 Off Low 10
High 100
BT/BT* 25.0 to 43.0
o
C (77.0 to 109.4
o
F) Off Low 34.0
o
C (93.2
o
F)
High 39.0
o
C (102.2
o
F)
CPP/CPP2/
CPP3/CPP4
-25 to 300 mmHg (-3.3 to 40.0 kPa) Off Low 70 mmHg (9.3 kPa)
High 100 mmHg (13.3 kPa)
CVP -5 to 300 mmHg (-0.7 to 40.0 kPa) Off Low 0 mmHg (0.0 kPa)
High 20 mmHg (2.7 kPa)
et DES/i DES 0 to 20 % Off Low 0% High 20%

Table of Contents

Other manuals for Dräger Infinity Kappa

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Dräger Infinity Kappa and is the answer not in the manual?

Dräger Infinity Kappa Specifications

General IconGeneral
BrandDräger
ModelInfinity Kappa
CategoryMedical Equipment
LanguageEnglish

Related product manuals