ELECTROMAGNETIC COMPATIBILITY (EMC)
VF8 DELTA/DELTA XL/KAPPA B-7
Recommended separation distances
Recommended separation distances between portable and mobile RF communications
equipment and the equipment
Rated
maximum
output
power of
transmitter
W
Separation distance according to frequency of transmitter in meters
150 kHz - 80 MHz 80 MHz to 800MHz 800 MHz to 2.5 GHz
0.01 0.12 0.12 0.23
0.1 0.38 0.38 0.73
1 1.2 1.2 2.3
10 3.8 3.8 7.3
100 12 12 23
NOTE:
• For transmitters rated at a maximum output power not listed above, the recommended separation
distance d in meters (m) can be estimated using the equation applicable to the frequency of the
transmitter, where P is the maximum output power rating of the transmitter in watts (W) according
to the transmitter manufacturer.
• At 80 MHz and 800 MHz, the separation distance for the higher frequency range applies.
• These guidelines may not apply in all situations. Electromagnetic propagation is affected by
absorption and reflection from structures, objects and people.
d1.2
P= d1.2
P=
d 2.3
P=