WIRELESS N ETWORK
VF8 DELTA/DELTA XL/KAPPA 3-7
Wireless Network Safety Considerations
When operating the monitor in a wireless network, observe the following:
While the monitor is transmitting or receiving signals, do not hold the
transmitting/receiving unit close to exposed body parts, especially the face or
eyes. The antenna/wireless card should be at least 5 cm (2 inch) away from
the body.
Operation of the wireless network relies on uninterrupted signal transmission
between the transmitting and receiving components of the network. When
using the wireless network, be aware that
certain structural limitations within the hospital building may interfere with
signal transmission,
other devices emitting radio frequencies, such as leaky microwave ovens or
warmers, may interfere with signal transmission,
the frequencies emitted by the device may interfere with the operation of
other medical equipment.
Wireless mode does not support Infinity Explorer.
The maximum number of wireless beds per dedicated access point is six.
Wireless Network Setup
NOTE: Wireless mode is available only if the monitor is in DirectNet mode (see page
3-12) or during Pick and Go (see page 3-5).

Table of Contents

Other manuals for Dräger Infinity Kappa

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Dräger Infinity Kappa and is the answer not in the manual?

Dräger Infinity Kappa Specifications

General IconGeneral
BrandDräger
ModelInfinity Kappa
CategoryMedical Equipment
LanguageEnglish

Related product manuals