EasyManuals Logo

Dräger Infinity Kappa User Manual

Dräger Infinity Kappa
614 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Page #554 background imageLoading...
Page #554 background image
B-28 DELTA/DELTA XL/KAPPA VF8
B T
ECHNICAL DATA
Trident (NMT) Pod
BISx Pod
Physical Attributes
Size (H x W x T) 36 x 132 x 140 mm (1.42 x 5.19 x 5.51 in)
Weight 595 g (21 oz.)
Connections PodCom: for connection to monitor
•NMT Cable
EEG: for connection to an EEG pod or BISx pod.
Electric shock
protection:
Type CF
Mode of operation:
Continuous
Standards: IEC 60601-1
IEC 60601-2-40: Particular Requirements for Safety - Specification for
Electromyographs and Evoked Response Equipment.
Defibrillator proof?
No
Environmental Requirements
Temperature range:
Operating:10 to 40 °C (50 to 104 °F)
Storage: –20 to 40 °C (–4 to +104 °F)
Relative humidity:
Operating:20 to 90 %, non-condensing
Storage: 20 to 90 %, (with packaging)
Atmospheric pressure
Operating:525 to 795.1 mmHg (70 to 106 kPa)
Storage: 375 to 795.1 mmHg (50 to 106 kPa)
Physical Attributes
Size (H x W x T) 114.3 x 95.3 x 63.5 mm (4.5 x 3.75 x 2.5 in)
Weight (without
PodCom cable)
227 g (8 oz.)
Connections PodCom: for connection to monitor
Power source:
Powered directly from monitor
Protection against
electric shock:
Type BF
Mode of operation:
Continuous

Table of Contents

Other manuals for Dräger Infinity Kappa

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Dräger Infinity Kappa and is the answer not in the manual?

Dräger Infinity Kappa Specifications

General IconGeneral
BrandDräger
ModelInfinity Kappa
CategoryMedical Equipment
LanguageEnglish

Related product manuals