10-6 DELTA/DELTA XL/KAPPA VF8
10 ST M
ONITORING
Measuring Points
The Measuring Points Screen
The starting and ending points for the QRS complex are automatically determined. In
practice, however, the accurate determination of the isoelectric and ST measuring
points requires careful clinical evaluation. On the Measuring Points screen, accessible
via the ST Analysis menu (page 10-4), you can change the isoelectric and ST
measuring points to achieve an accurate ST deviation measurement.
Measuring
Points
Displays the average S-T complex for
each monitored ST lead, as well as
TruST leads.
NOTE: See page 10-6 for more
information.
ISO; changes the point that defines
position of the isoelectric point.
ST;
changes the point that defines
position of the S-T measurement
point.
Recalculates the QRS complex.
ST Alarms Opens the ST Alarm Limits table. See page 10-10.