NBP SETUP
VF8 DELTA/DELTA XL/KAPPA 13-3
NBP Setup
Safety Considerations
Software and Hardware Cuff Pressure Cutoffs
The cuff deflates automatically if a measurement takes longer than 2 minutes in Adult/
Pediatric
mode or 90 seconds in Neonatal mode. To protect the patient from
excessive pressure, inflation limits have been established in all patient categories; see
Appendix B, for more information.
Cuff Selection and Placement
The quality of NBP monitoring depends largely on the quality of the signals received
by the monitor. It is important to select the correct cuff size for your patient. Measure
the circumference of your patient’s limb and use the markings on the cuff to select the
correct cuff size.
WARNING: Obstructions may cause the cuff to inflate and
deflate improperly and result in inaccurate readings. Check
the hose and cuff for damage and dirt. Do not allow the hose
and cuff to come in contact with fluids, and make sure that
they are not compressed or kinked.
WARNING: Accurate NBP measurements depend on the
correct size and type of the blood pressure cuff in relation to
the patient's arm circumference. The wrong sized cuff, or
cuffs outside the range or size manufactured by Dräger, can
cause inaccurate measurements. Use only Dräger approved
cuffs and make sure that the correct size is used for each
patient.