INFINITY DELTA SERIES USER’S GUIDE
VF8 DELTA/DELTA XL/KAPPA 7
Site of Operation
Only use these devices in areas that meet the environmental requirements outlined in
the technical data section.
WARNING:
Do not operate the device in areas such as: magnetic
resonance imaging (MRI) environments, aircraft,
ambulance, home or hyperbaric chambers.
Do not operate devices (monitor, pods, modules and
accessories) in close proximity to equipment that emits
microwave or other high-frequency emissions since
they may interfere with the devices’ operation.
When placing the device make sure adequate
ventilation exists and prevent overheating by
positioning this device with at least 2 in (5 cm) of space
around all sides. Do not cover the devices with
blankets or bedsheets. To prevent burns to the patient
avoid direct contact between these items’ external
surfaces and the patient.
Only the items indicated on the list of accessories in
the “Approved Options and Accessories” chapter have
been tested and approved to be used with the device.
Accordingly it is strongly recommended that only these
accessories be used in conjunction with the specific
device. Otherwise the correct functioning of the device
may be compromised.
Disposable accessories (such as disposable
electrodes, transducers, etc.) are for single use only.
Do not reuse disposable accessories.
To minimize the risk of patient strangulation, carefully
position and secure sensor cables. Also carefully
position sensor cables to minimize inductive loops.
To avoid explosions, devices should not be used in the
presence of flammable anesthetic mixture including
oxygen, ether, nitrous oxide, and cyclopropane.
Because of the danger of electric shock, never remove
the cover of a device while it is in operation or
connected to power.