MICRO2+® STANDALONE PULSE OXIMETER
VF8 DELTA/DELTA XL/KAPPA 17-21
Parameters
SpO
2
MicrO2+ Setup
To access the SpO
2
MicrO2+ menu
Click on the SpO
2
* parameter box
or
1. Press the
Menu fixed key.
2. Click on
Patient Setup.
3. Click on
Parameters to display a list of available parameters.
4. Click on
SpO2*.
Click on
SpO2* Label to set the label in the parameter box. Choices are None, Pre-
ductal,
and Post-ductal. If you choose None, the SpO2* parameter box will have no
label.
Click on
SpO2* Alarm to go to the SpO2* entry in the Alarm Limits menu.
Parameter Label Units Range
SpO
2
SpO
2
* % 1 to 100
PLS PLS* b/min 30 to 250
Delta SpO
2
SpO
2
% % 0 to 99
NOTE:
SpO
2
% is the absolute value of (SpO
2
- SpO
2
*).
Both SpO
2
and SpO
2
* must be connected to get a SpO
2
% value.
If PLS and PLS* are not within +/- 6 bpm, the SpO
2
% parameter field will be
blank.

Table of Contents

Other manuals for Dräger Infinity Kappa

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Dräger Infinity Kappa and is the answer not in the manual?

Dräger Infinity Kappa Specifications

General IconGeneral
BrandDräger
ModelInfinity Kappa
CategoryMedical Equipment
LanguageEnglish

Related product manuals