15-14 DELTA/DELTA XL/KAPPA VF8
15 C
ARDIAC OUTPUT (C.O.)
Status Messages
Message Possible Cause Suggested Action
BT > UL
BT < LL
The blood temperature is outside the
alarm limits, due to one of the
following conditions:
A physiological condition
Inappropriate alarm limits
A defective sensor or cartridge
Check the patient and treat, if
necessary.
Change the alarm limits.
Check the equipment and replace
it, if necessary.
BT Out of Range
(High)
BT Out of Range
(Low)
The blood temperature is outside the
measurement range (25 °C to 43 °C),
because of a defective sensor or
cartridge.
Check the equipment and replace it,
if necessary.
C.O. Already In Use The catheter and injectate probe are
connected to different hemo pods.
Disconnect the intermediate
cables from the hemo pods.
Connect both the catheter and the
injectate probe to the same cable
and reconnect them.
Discard the unused intermediate
cable.
C.O. Out of Range
(High)
C.O. Out of Range
(Low)
The cardiac output value is greater
than 20 liters/min or less than 0.5 liter/
min because of the following
conditions:
A physiological condition
Unstable baseline
Incorrect injectate volume, catheter
size, or Comp. Constant
Defective catheter, cable, or
cartridge
Check the patient and treat, if
necessary.
Use cooler injectate.
Enter the correct values in the
C.O. menu.
Repeat the measurement. If the
message persists, replace the
defective components.
C.O. Injectate Too
Cold
The injectate is cooler than –5 °C. Use an injectate within the range of
–5 °C to +30 °C.
Defective cable or hemodynamic pod. Check the equipment and replace it,
if necessary.
C.O. Injectate Too
Warm
The injectate is warmer than +30 °C. Use an injectate within the range of
–5 °C to +30 °C.
The injectate probe is not connected. Check probe connection. If problem
persists, replace the probe.
Defective cable or hemodynamic pod. Repeat the measurement. If the
problem persists, replace defective
part.

Table of Contents

Other manuals for Dräger Infinity Kappa

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Dräger Infinity Kappa and is the answer not in the manual?

Dräger Infinity Kappa Specifications

General IconGeneral
BrandDräger
ModelInfinity Kappa
CategoryMedical Equipment
LanguageEnglish

Related product manuals