MONITORING A CCESSORIES
VF8 DELTA/DELTA XL/KAPPA B-21
Monitoring Accessories
etCO
2
Module
Physical Attributes
Size (H x W x D):
Module: 150 x 93 x 65 mm (5.9 x 3.6 x 2.6 in)
Capnostat III sensor:
33 x 42 x 22 mm (1.3 x 1.7 x 0.9 in)
Weight:
Module: 0.5 kg (1.1 lb)
Capnostat
TM
III sensor:18 g
Connections:
Sensor connector, female luer side-stream sampling port, male luer
sample exhaust port
Adult airway adapter dead
space:
<5 cc
Neonatal airway adapter
dead space:
< 0.5 cc
Moisture resistance:
Airway adapter can be immersed in water without damage
NOTE: Capnostat
TM
III sensor size and weight exclude cable.
Electrical Specifications
Power source:
Powered directly from monitor
Protection against electric
shock:
Type CF
Mode of operation:
Continuous
Environmental Requirements
Temperature range:
Operating:10 to 40 °C (50 to 104 °F)
Storage: –20 to 50 °C (–4 to 122 °F)
Relative humidity:
Operating:20 to 90 %, non-condensing
Storage: 10 to 95 % (with packaging)
Atmospheric pressure:
Operating:525 to 795 mmHg (70 to 106 kPa)
Storage: 375 to 795 mmHg (50 to 106 kPa)