STATUS M ESSAGES
VF8 DELTA/DELTA XL/KAPPA 9-9
Status Messages
Archive Determines whether the selected event is
stored, recorded, or both. You can view
stored events on the Event Recall screen
(see page 1-27).
NOTE: The archive function for ASY
(Asystole) and VF (Ventricular Fibrillation)
cannot be disabled.
• Store; stores the selected event
• Record; generates an alarm recording
of selected event
• Str./Rec.; stores an event and
generates an alarm recording (even
when event alarm is turned off)
•OFF
Message Event Definition Suggested Action
PVC/min > UL The PVC/min value is above the upper alarm
limit.
Check the patient.
Cannot Learn
Lead <X>
At the end of the Learning phase, the
dominant normal complex could not be
determined for Lead <X> (one of the two
leads selected for QRS processing).
Arrhythmia analysis proceeds using the other
lead as a source. Lead <X> is ignored until a
Relearn is initiated.
• Check the leads.
• Choose another lead for QRS
processing.
Relearning Monitor is learning a normal QRS complex as
a reference template.
Not applicable
Unable to
Learn
After 100 beats, monitor cannot determine
the dominant normal complex on any lead
selected for QRS processing. Learning
continues.
Check the electrode preparation.
Baseline
Artifact
Artifact is blocking arrhythmia classification. Check the electrode preparation.
Function Description Available Settings