IBP
VF8 DELTA/DELTA XL/KAPPA 29-5
PiCCO Pod
1. Disconnect the pod from the monitor.
2. Clean the pod with a gauze moistened in enzymatic detergent or a solution of
green tinctured soap and water. Dry thoroughly with a lint-free cloth.
MPod – Quad Hemo
Agents tested by Dräger and shown to have no harmful effect at the time of testing on
the materials utilized in the MPod – Quad Hemo include:
Isopropyl alcohol 40 %
Compliance™; this cleaning agent may discolor soft plastic material
Sporox II
Dismozon pur
CAUTION:
Do not use organic solvents.
Do not sterilize by steam, heat, radiation or ETO.
Do not use sharp objects.
Make sure that NO liquid enters the pod.
CAUTION: Do not autoclave the MPod - Quad Hemo.
CAUTION: If using alcohol, it should only be a 40 % diluted solution. Higher
concentrations could damage the device.