Chapter 2 105
Adjustment Procedures
13. Demodulator Adjustment
13. Demodulator Adjustment
Assembly Adjusted
A4 log amplifier assembly
Related Performance Test
There is no related performance test for this adjustment procedure.
Description
A 5 kHz peak deviation FM signal is applied to the INPUT 50. The
detected audio is monitored by an oscilloscope. FM DEMOD is adjusted
to peak the response displayed on the oscilloscope.
Figure 2-18 Demodulator Adjustment Setup