42 Chapter1
General Information
Electrostatic Discharge
Reducing Potential for ESD Damage
The suggestions that follow may help reduce ESD damage that occurs
during instrument testing and servicing:
Before connecting any coaxial cable to a spectrum analyzer
connector for the first time each day, momentarily ground the center
and outer connectors of the cable.
Personnel should be grounded with a resistor-isolated wrist strap
before touching the center in of any connector and before removing
any assembly from the unit.
Be sure all instruments are properly earth-grounded to prevent
build-up of static discharge.
Static-Safe Accessories
Table 1-4 Static-Safe Accessories
HP Part
Number
Description
9300-0797 Set includes: 3M static control mat 0.6 m × 1.2m(2
ft × 4 ft) and 4.6 cm (15 ft) ground wire. (The
wrist-strap and wrist-strap cord are not included.
They must be ordered separately.)
9300-0980 Wrist-strap cord 1.5 m (5 ft)
9300-1383 Wrist-strap, color black, stainless steel, without
cord, has four adjustable links and a 7 mm
post-type connection.
9300-1169 ESD heel-strap (reusable 6 to 12 months).