EasyManuals Logo
Home>Keithley>Test Equipment>2450

Keithley 2450 User Manual

Keithley 2450
878 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Page #148 background imageLoading...
Page #148 background image
Model 2450
Interactive SourceMeter® Instrument Reference Manual Section 3:
Functions and features
2450-901-01 Rev. B/September 2013 3-3
To set the relative offset for another function, replace VOLTage with CURRent or RESistance.
Using TSP commands:
Send the commands:
smu.measure.func = smu.FUNC_DC_VOLTAGE
smu.measure.rel.level = relValue
smu.measure.rel.enable = smu.ON
Where relValue is the relative offset value.
To set the relative offset for another function, replace smu.FUNC_DC_VOLTAGE with
smu.FUNC_DC_CURRENT or smu.FUNC_RESISTANCE.
Disabling the relative offset
Using the front panel:
1. Select the measurement function to which the relative offset is applied.
2. On the SETTINGS swipe screen, select the box next to Rel. An X should be displayed and the
REL annunciator to the right of the measurement is no longer displayed.
The relative offset feature can also be disabled by selecting MENU > Filter/Math > Rel State and
selecting Off.
Using SCPI commands:
Send the command:
:SENSe:VOLTage:RELative OFF
To set the relative offset for another function, replace VOLTage with CURRent or RESistance.
Using TSP commands:
Send the commands:
smu.measure.func = smu.FUNC_DC_VOLTAGE
smu.measure.rel.enable = smu.OFF
To set the relative offset for another function, replace smu.FUNC_DC_VOLTAGE with
smu.FUNC_DC_CURRENT or smu.FUNC_RESISTANCE.

Table of Contents

Other manuals for Keithley 2450

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Keithley 2450 and is the answer not in the manual?

Keithley 2450 Specifications

General IconGeneral
BrandKeithley
Model2450
CategoryTest Equipment
LanguageEnglish

Related product manuals