Interactive SourceMeter® Instrument Reference Manual Section 8:
2450-901-01 Rev. B/September 2013 8-243
tspnet.reset()
This function disconnects all TSP-Net sessions.
Type TSP-Link accessible
Affected by Where saved Default value
Usage
tspnet.reset()
Details
This command disconnects all remote instruments connected through TSP-Net. For TSP-enabled devices, this
causes any commands or scripts running remotely to be terminated.
Also see
None
tspnet.termination()
This function sets the device line termination sequence.
Type TSP-Link accessible
Affected by Where saved Default value
Usage
type = tspnet.termination(connectionID)
type = tspnet.termination(connectionID, termSequence)
• tspnet.TERM_LF
• tspnet.TERM_CR
• tspnet.TERM_CRLF
• tspnet.TERM_LFCR
Details
This function sets and gets the termination character sequence that is used to indicate the end of a line for a
TSP-Net connection.
Using the termSequence parameter sets the termination sequence. The present termination sequence is
always returned.
For the termSequence parameter, use the same values listed in the table above for type. There are four
possible combinations, all of which are made up of line feeds (LF or 0x10) and carriage returns (CR or 0x13). For
TSP-enabled devices, the default is tspnet.TERM_LF. For devices that are not TSP-enabled, the default is
tspnet.TERM_CRLF.