6: SCPI command reference Model 2450 Interactive SourceMeter® Instrument
6-18 2450-901-01 Rev. B/September 2013
:CALCulate2:<function>:LIMit<Y>:UPPer[:DATA]
This command specifies the upper limit for a limit test.
Type Affected by Where saved Default value
Command and query
Recall settings
Instrument reset
Power cycle
Measure configuration list
Save settings
Measure configuration list
1.000000E+00
Usage
:CALCulate2:<function>:LIMit<Y>:UPPer[:DATA] <n>
:CALCulate2:<function>:LIMit<Y>:UPPer[:DATA]?
:CALCulate2:<function>:LIMit<Y>:UPPer[:DATA]? DEFault
:CALCulate2:<function>:LIMit<Y>:UPPer[:DATA]? MINimum
:CALCulate2:<function>:LIMit<Y>:UPPer[:DATA]? MAXimum
The function to which this setting applies:
• Current: CURRent[:DC]
• Resistance: RESistance
•
The limit test that the upper limit applies to, 1 or 2
The value of the upper limit (
−
9.99999e+11 to +9.99999e+11)
Details
This command sets the high limits for the limit tests for the selected measurement function. When limit testing is
enabled for this limit, the instrument generates a fail indication when the measurement value is more than this
value.
If you send this command without the <function> parameter, it will set the limit for all functions.
Example
CALC2:VOLT:LIM1:UPP 2.5
CALC2:VOLT:LIMIT1:STAT ON
Set lower limit 1 for voltage to 0.25 volts.
Set upper limit 1 for voltage to 2.5 volts.
Enable limit 1 testing for voltage.
Also see
:CALCulate2:<function>:LIMit<Y>:LOWer[:DATA] (on page 6-16)