EasyManuals Logo
Home>Keithley>Test Equipment>2450

Keithley 2450 User Manual

Keithley 2450
878 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Page #359 background imageLoading...
Page #359 background image
Section
6: SCPI command reference Model 2450 Interactive SourceMeter® Instrument
Reference Manual
6-42 2450-901-01 Rev. B/September 2013
[:SENSe[1]]:COUNt
This command sets the number of measurements to be performed when a measurement is requested.
Type Affected by Where saved Default value
Command and query
Recall settings
Instrument reset
Power cycle
Save settings 1
Usage
[:SENSe[1]]:COUNt <n>
[:SENSe[1]]:COUNt DEFault
[:SENSe[1]]:COUNt MINimum
[:SENSe[1]]:COUNt MAXimum
[:SENSe[1]]:COUNt?
[:SENSe[1]]:COUNt? DEFault
[:SENSe[1]]:COUNt? MINimum
[:SENSe[1]]:COUNt? MAXimum
<n>
The number of measurements (1 to 300,000)
Details
This command sets the number of measurements that are taken when a measurement is requested. This
command does not affect the trigger model.
To get better feedback from the instrument, use the Simple Loop trigger model template instead of
using the count command.
Example
COUN 10
Make ten measurements.
Also see
:MEASure:<function>? (on page 6-4)

Table of Contents

Other manuals for Keithley 2450

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Keithley 2450 and is the answer not in the manual?

Keithley 2450 Specifications

General IconGeneral
BrandKeithley
Model2450
CategoryTest Equipment
LanguageEnglish

Related product manuals