8: TSP command reference Model 2450 Interactive SourceMeter® Instrument
Reference Manual
8-166 2450-901-01 Rev. B/September 2013
trigger.blender[N].clear()
This function clears the blender event detector and resets the overrun indicator of blender N.
Type TSP-Link accessible
Affected by Where saved Default value
Function
Yes
Usage
trigger.blender[N].clear()
N
The blender number (1 or 2)
Details
This command sets the blender event detector to the undetected state and resets the overrun indicator of the
event detector.
Example
trigger.blender[2].clear()
Clears the event detector for
blender 2.
Also see
None
trigger.blender[N].orenable
This attribute selects whether the blender performs OR operations or AND operations.
Type TSP-Link accessible Affected by Where saved Default value
Attribute (RW)
Yes
Restore configuration
Instrument reset
Power cycle
Trigger blender N reset
Configuration script
false (AND)
Usage
orenable = trigger.blender[N].orenable
trigger.blender[N].orenable = orenable
orenable
The type of operation:
true: OR operation
false
: AND operation
N
The trigger blender (1 or 2)
Details
This command selects whether the blender waits for any one event (OR) or waits for all selected events (AND)
before signaling an output event.

Table of Contents

Other manuals for Keithley 2450

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Keithley 2450 and is the answer not in the manual?

Keithley 2450 Specifications

General IconGeneral
BrandKeithley
Model2450
CategoryTest Equipment
LanguageEnglish

Related product manuals