Model 2450
Interactive SourceMeter® Instrument Reference Manual Section 6:
SCPI command reference
2450-901-01 Rev. B/September 2013 6-57
[:SENSe[1]]:<function>:RELative:STATe
This command enables or disables the relative offset value.
Type Affected by Where saved Default value
Command and query
Recall settings
Instrument reset
Power cycle
Measure configuration list
Save settings
Measure configuration list
0 (OFF)
Usage
[:SENSe[1]]:<function>:RELative:STATe <b>
[:SENSe[1]]:<function>:RELative:STATe?
<function>
The measurement function to which this setting applies:
Current: CURRent[:DC]
Resistance: RESistance
Voltage:
VOLTage[:DC]
<b>
Disable the relative offset: 0 or OFF
Enable the relative offset: 1 or ON
Details
This attribute enables or disables relative measurements. When relative measurements are enabled, all
subsequent measured readings are offset by the relative offset value calculated when you acquire the relative
offset value.
Each returned measured relative reading is the result of the following calculation:
Display value = Actual measured value - Relative offset value
Example
SENS:FUNC "VOLT"
SENS:VOLT:REL 5
:SENSe:VOLT:REL:STATe ON
Set the measurement function to volts with a
relative offset of 5 V and enable the relative
offset function.
Also see
[:SENSe[1]]:<function>:RELative (on page 6-55)
[:SENSe[1]]:<function>:RELative:ACQuire (on page 6-56)

Table of Contents

Other manuals for Keithley 2450

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Keithley 2450 and is the answer not in the manual?

Keithley 2450 Specifications

General IconGeneral
BrandKeithley
Model2450
CategoryTest Equipment
LanguageEnglish

Related product manuals