Interactive SourceMeter® Instrument Reference Manual Section 8:
2450-901-01 Rev. B/September 2013 8-135
smu.source.highc
This attribute enables or disables high capacitance feature.
Type TSP-Link accessible Affected by Where saved Default value
Attribute (RW) Yes
Restore configuration
Instrument reset
Power cycle
Source configuration list
Configuration script
Source configuration list
smu.OFF
Usage
state = smu.source.highc
smu.source.highc = state
Turn high capacitance off:
or
Turn high capacitance on: smu.ON or 1
Details
When the instrument is measuring low current and is driving a capacitive load, you may see overshoot, ringing,
and instability. You can enable the high capacitance feature to minimize these problems.
High-capacitance settings apply when operating using the 10 nA through the 100 mA current ranges. When
operating using the 1 A range, the high-capacitance setting will not affect the instrument rise time or current
measure settling time.
Example
smu.source.highc = smu.ON
Turn the high capacitance feature on.
Also see
High capacitance operation (on page 4-21)
smu.source.level
This attribute immediately selects a fixed amplitude for the selected source function.
Type TSP-Link accessible Affected by Where saved Default value
Instrument reset
Power cycle
Source configuration list
Function change
Source configuration list
Usage
sourceLevel = smu.source.level
smu.source.level = sourceLevel
Current: -1.05 A to 1.05 A
Voltage: -210 V to 210 V