8: TSP command reference Model 2450 Interactive SourceMeter® Instrument
Reference Manual
8-164 2450-901-01 Rev. B/September 2013
status.standard.event
This attribute returns the contents of the Standard Event Status Register set of the status model.
Type TSP-Link accessible Affected by Where saved Default value
Attribute (R)
Yes
Status preset()
Not saved
0
Usage
standardRegister = status.standard.event
standardRegister
The status of the standard event status register
Details
When this command returns zero (0), no bits are set. You can send 0 to clear all bits.
The instrument returns a decimal value that corresponds to the binary-weighted sum of all bits set in the register.
Bit
Decimal
value
Constant When set, indicates the following has occurred:
0
1
status.standard.OPC
All pending selected instrument operations are
complete and the instrument is ready to accept new
commands. The bit is set in response to an
*OPC (on
page B-6) command or TSP opc() (on page 8-77)
function.
1
2
Not used
Not used.
2 4
status.standard.QYE
Attempt to read data from an empty Output Queue.
3 8
status.standard.DDE
An instrument operation did not execute properly due
to an internal condition.
4
16
status.standard.EXE
The instrument detected an error while trying to
execute a command.
5 32
status.standard.CME
A command error has occurred. See information
following this table for descriptions of command errors.
6
64
status.standard.URQ
The instrument transitioned from remote control to
local control.
7 128
status.standard.PON
The instrument has been turned off and turned back on
since the last time this register was read.
Command errors include:
IEEE Std 488.2 syntax error: The instrument received a message that does not follow the defined
syntax of the IEEE Std 488.2 standard.
Semantic error: The instrument received a command that was misspelled or received an optional IEEE
Std 488.2 command that is not implemented in the instrument.
GET error: The instrument received a Group Execute Trigger (GET) inside a program message.
Example
print(status.standard.event)
May return the value 149, showing that the Standard Event
Status Register contains binary
10010101
Also see
Standard Event Register (on page C-3)
Understanding bit settings (on page C-15)

Table of Contents

Other manuals for Keithley 2450

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Keithley 2450 and is the answer not in the manual?

Keithley 2450 Specifications

General IconGeneral
BrandKeithley
Model2450
CategoryTest Equipment
LanguageEnglish

Related product manuals