Model 2450
Interactive SourceMeter® Instrument Reference Manual Section 6:
SCPI command reference
2450-901-01 Rev. B/September 2013 6-29
:FORMat:BORDer
This command sets the byte order for the IEEE-754 binary formats.
Type Affected by Where saved Default value
Command and query
Recall settings
Instrument reset
Power cycle
Save settings SWAP
Usage
:FORMat:BORDer <name>
:FORMat:BORDer?
<name>
The binary byte order:
Normal byte order: NORMal
Reverse byte order for binary formats: SWAPped
Details
This attribute selected the byte order in which data is written.
The SWAPped byte order must be used when transmitting binary data to a computer with a Microsoft Windows
operating system.
The ASCII data format can only be sent in the normal byte order. If the ASCII format is selected, the SWAPped
selection is ignored.
When you select NORMal byte order, the data format for each element is sent as follows:
Byte 1 Byte 2 Byte 3 Byte 4 (Single precision)
When you select SWAPped, the data format for each element is sent as follows:
Byte 4 Byte 3 Byte 2 Byte 1 (Single precision)
The “#0” header is not affected by this command. The header is always sent at the beginning of the data string
for each measurement conversion.
Example
FORM:BORD NORM
Use the normal byte order.
Also see
:FORMat[:DATA] (on page 6-30)

Table of Contents

Other manuals for Keithley 2450

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Keithley 2450 and is the answer not in the manual?

Keithley 2450 Specifications

General IconGeneral
BrandKeithley
Model2450
CategoryTest Equipment
LanguageEnglish

Related product manuals