6: SCPI command reference Model 2450 Interactive SourceMeter® Instrument
6-40 2450-901-01 Rev. B/September 2013
[:SENSe[1]]:CONFiguration:LIST:RECall
This command recalls a specific configuration point in a specific measure configuration list.
Type Affected by Where saved Default value
Usage
[:SENSe[1]]:CONFiguration:LIST:RECall <name>
[:SENSe[1]]:CONFiguration:LIST:RECall <name>, <point>
A string that represents the name of a measure configuration list
A specific configuration point in the configuration list
Details
Use this command to recall the settings stored in a specific configuration point on a specific configuration list. If
you do not specify a point when you send the command, it recalls the settings stored in the first configuration
point on the specified configuration list.
Examples
:SENSe:CONF:LIST:RECall "MyMeasList", 5
Recalls configuration point 5 in a configuration list
named MyMeasList.
:SENSe:CONF:LIST:RECall "MyMeasList"
Since a point was not specified, this command
recalls configuration point 1 from a configuration
list named
.
Also see
Configuration lists (on page 3-33)
[:SENSe[1]]:CONFiguration:LIST:CREate (on page 6-37)
[:SENSe[1]]:CONFiguration:LIST:STORe (on page 6-41)
[:SENSe[1]]:CONFiguration:LIST:SIZE?
This command returns the size (number of configuration points) of a measure configuration list.
Type Affected by Where saved Default value
Usage
[:SENSe[1]]:CONFiguration:LIST:SIZE? <name>
A string that represents the name of a measure configuration list