Model 2450
Interactive SourceMeter® Instrument Reference Manual Section 8:
TSP command reference
2450-901-01 Rev. B/September 2013 8-245
tspnet.tsp.abort()
This function causes the TSP-enabled instrument to stop executing any of the commands that were sent to it.
Type TSP-Link accessible
Affected by Where saved Default value
Function
No
Usage
tspnet.tsp.abort(connectionID)
connectionID
Integer value used as a handle for other
tspnet
commands
Details
This function is appropriate only for TSP-enabled instruments.
Sends an abort command to the remote instrument.
Example
tspnet.tsp.abort(testConnection)
Stops remote instrument execution on testConnection.
Also see
None
tspnet.tsp.abortonconnect
This attribute contains the setting for abort on connect to a TSP-enabled instrument.
Type TSP-Link accessible Affected by Where saved Default value
Attribute (RW)
No
Restore configuration
Instrument reset
Power cycle
Configuration script
1 (enable)
Usage
tspnet.tsp.abortonconnect = value
value = tspnet.tsp.abortonconnect
value
1 (enable) or 0 (disable)

Table of Contents

Other manuals for Keithley 2450

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Keithley 2450 and is the answer not in the manual?

Keithley 2450 Specifications

General IconGeneral
BrandKeithley
Model2450
CategoryTest Equipment
LanguageEnglish

Related product manuals