8: TSP command reference Model 2450 Interactive SourceMeter® Instrument
8-74 2450-901-01 Rev. B/September 2013
Details
Enable this attribute to have the instrument automatically send generated events that are stored in the event log.
The event log is cleared when the event is sent.
Events are processed after a command message is executed but before prompts are issued (if prompts are
enabled with localnode.prompts).
If this attribute is disabled, errors are left in the event log and must be explicitly read or cleared.
If you are using this command from a remote node, replace localnode with the node reference, such as
node[5].showevents.
Example
Send generated warning and information
messages, but do not send error messages.
Also see
eventlog.clear() (on page 8-54)
localnode.prompts (on page 8-71)
localnode.version
This attribute stores the instrument version.
Type TSP-Link accessible Affected by Where saved Default value
Usage
version = localnode.version
Details
This attribute indicates the version number of the firmware that is presently running in the instrument.
When using this command from a remote node, localnode should be replaced with the node reference. For
example, node[5].version.
Example
Outputs the present version level. Example output:
Also see
localnode.model (on page 8-70)
localnode.serialno (on page 8-72)