Section
2: General operation Model 2450 Interactive SourceMeter® Instrument
Reference Manual
2-4 2450-901-01 Rev. B/September 2013
Force terminals
Use FORCE HI and FORCE LO terminal connections to source
or sink voltage or current to or from a device under test (DUT).
FRONT/REAR
TERMINALS switch
Press the FRONT/REAR TERMINALS switch to activate
terminals on the front or rear panel. When the front-panel
terminals are active, a green "F" is visible to the left of the
FRONT/REAR switch. When the rear-
panel terminals are active, a
yellow "R" is visible to the left of the switch.
Chassis
connection
Banana jack connector that provides a chassis connection.
Power the instrument on or off
Follow the procedure below to connect the Model 2450 to line power and turn on the instrument. The
Model 2450 operates from a line voltage of 100 V to 240 V at a frequency of 50 Hz or 60 Hz. Line
voltage is automatically sensed. Make sure the operating voltage in your area is compatible.
The Model 2450 must be turned on and allowed to warm up for at least one hour to achieve rated
accuracies.
Operating the instrument on an incorrect line voltage may cause damage to the instrument, possibly
voiding the warranty.
The power cord supplied with the Model 2450 contains a separate protective earth (safety
ground) wire for use with grounded outlets. When proper connections are made, the
instrument chassis is connected to power-line ground through the ground wire in the power
cord. In addition, a redundant protective earth connection is provided through a screw on
the rear panel. This terminal should be connected to a known protective earth. In the event
of a failure, not using a properly grounded protective earth and grounded outlet may result
in personal injury or death due to electric shock.

Table of Contents

Other manuals for Keithley 2450

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Keithley 2450 and is the answer not in the manual?

Keithley 2450 Specifications

General IconGeneral
BrandKeithley
Model2450
CategoryTest Equipment
LanguageEnglish

Related product manuals