Model 2450
Interactive SourceMeter® Instrument Reference Manual Section 6: SCPI command
reference
2450-901-01 Rev. B/September 2013 6-95
SYSTem subsystem
This subsystem contains commands that affect the overall operation of the instrument, such as
access, beepers, communications, event logs, and time.
:SYSTem:ACCess
This command contains the type of access users have to the instrument through different interfaces.
Type Affected by Where saved Default value
Command and query
Not applicable
Nonvolatile memory
FULL
Usage
:SYSTem:ACCess <permissions>
:SYSTem:ACCess?
<permissions>
The level of access that is allowed:
Full access for all users from all interfaces: FULL
Allows access by one remote interface at a time with login and logout
required from other interfaces: EXCLusive
Allows access by one remote interface at a time with passwords required on
all interfaces: PROTected
Allows access by one interface at a time (including the front panel) with
passwords required on all interfaces: LOCKout
Details
When access is set to full, the instrument accepts commands from any interface with no passwords required.
When access is set to exclusive, you must log out of one remote interface and log into another one to change
interfaces. To use another interface, log out of the present interface before logging into the new interface. You do
not need a password with this access.
Protected access is similar to exclusive access, except that you must enter a password when logging in.
When the access is set to locked out, a password is required to change interfaces, including the front panel
interface.
Under any access type, if a script is running on one interface when a command comes in from another interface,
the command is ignored and the message "Script running from another interface" occurs.
The command *idn? is permitted from any interface in all access types.
Example
:SYST:ACC LOCK
login admin
logout
Set the instrument access to locked out.
Log into the interface using the default password.
Log out of the interface.
Also see
:SYSTem:PASSword:NEW (on page 6-105)

Table of Contents

Other manuals for Keithley 2450

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Keithley 2450 and is the answer not in the manual?

Keithley 2450 Specifications

General IconGeneral
BrandKeithley
Model2450
CategoryTest Equipment
LanguageEnglish

Related product manuals