Section
8: TSP command reference Model 2450 Interactive SourceMeter® Instrument
Reference Manual
8-224 2450-901-01 Rev. B/September 2013
trigger.timer[N].wait()
This function waits for a trigger.
Type TSP-Link accessible
Affected by Where saved Default value
Function
Yes
Usage
triggered = trigger.timer[N].wait(timeout)
triggered
Trigger detection indication
N
Trigger timer number (1 to 4)
timeout
Maximum amount of time in seconds to wait for the trigger
Details
If one or more trigger events were detected since the last time trigger.timer[N].wait() or
trigger.timer[N].clear() was called, this function returns immediately.
After waiting for a trigger with this function, the event detector is automatically reset and rearmed. This is true
regardless of the number of events detected.
Example
triggered = trigger.timer[3].wait(10)
print(triggered)
Waits up to 10 seconds for a trigger on timer 3.
If false is returned, no trigger was detected during
the 10-second timeout.
If true is returned, a trigger was detected.
Also see
trigger.timer[N].clear() (on page 8-216)
trigger.tsplinkin[N].clear()
This function clears the event detector for a trigger.
Type TSP-Link accessible
Affected by Where saved Default value
Function
Yes
Usage
trigger.tsplinkin[N].clear()
N
The trigger line (1 to 3)

Table of Contents

Other manuals for Keithley 2450

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Keithley 2450 and is the answer not in the manual?

Keithley 2450 Specifications

General IconGeneral
BrandKeithley
Model2450
CategoryTest Equipment
LanguageEnglish

Related product manuals