EasyManuals Logo
Home>Keithley>Test Equipment>2450

Keithley 2450 User Manual

Keithley 2450
878 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Page #188 background imageLoading...
Page #188 background image
Model 2450
Interactive SourceMeter® Instrument Reference Manual Section 3:
Functions and features
2450-901-01 Rev. B/September 2013 3-43
Using remote commands for configuration list operations
The following topics provide information to:
Create an example source configuration list named MySourceList.
Store four example configuration points on MySourceList.
View the contents of a specific configuration point.
Recall a configuration point from MySourceList.
The following table provides information about specific settings for each configuration point.
Settings Index 1 Index 2 Index3 Index 4
Function
Voltage
Voltage
Voltage
Voltage
AutoRange
Off
Off
Off
Off
Delay
0.001
0.001
0.001
0.001
AutoDelay
Off
Off
Off
Off
Source Level
2.0 V
3.0 V
4.0 V
5.0 V
ProtectLevel
PROTECT_40
PROTECT_40
PROTECT_40
PROTECT_40
ProtectTripped
Not tripped
Not tripped
Not tripped
Not tripped
OffMode
OFFMODE_
NORMAL
OFFMODE_
NORMAL
OFFMODE_
NORMAL
OFFMODE_
NORMAL
Range
20.0 V
20.0 V
20.0 V
20.0 V
LimitLevel
10.6 nA
10.6 nA
10.6 nA
10.6 nA
LimitTripped
Not tripped
Not tripped
Not tripped
Not tripped
HighC
Off
Off
Off
Off
Readback
On
On
On
On
UserDelay1-5 None None None None
The figures in this section include a graphic representation of MySourceList and the four
configuration points. For simplicity, the figures will only show two of the source settings listed in the
table (source range and source level).
This example creates a source configuration list; however, you can use the same process to create a
measure configuration list.
Creating a configuration list and giving it a name
Use one of the following methods to create the configuration list used in this example and give it a
name.
Using SCPI commands:
:SOURce:CONFiguration:LIST:CREate "MySourceList"
Using TSP commands:
smu.source.configlist.create("MySourceList")

Table of Contents

Other manuals for Keithley 2450

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Keithley 2450 and is the answer not in the manual?

Keithley 2450 Specifications

General IconGeneral
BrandKeithley
Model2450
CategoryTest Equipment
LanguageEnglish

Related product manuals