Interactive SourceMeter® Instrument Reference Manual Section 6:
SCPI command reference
2450-901-01 Rev. B/September 2013 6-89
STATus subsystem
The STATus subsystem controls the status registers of the Model 2450. For additional information on
the status model, see Status model (on page C-1
).
:STATus:OPERation:CONDition?
This command reads the Operation Event Register of the status model.
Type Affected by Where saved Default value
Query only
Not applicable
Not applicable
Not applicable
Usage
:STATus:OPERation:CONDition?
Details
This command reads the contents of the Operation Condition Register, which is one of the Operation Event
Registers.
For detail on interpreting the value of a register, see Understanding bit settings (on page C-15
).
Example
:STAT:OPER:COND?
Returns the contents of the Operation
Condition Register.
Also see
Operation Event Register (on page C-8)
:STATus:OPERation:ENABle
This command sets or reads the contents of the Operation Event Enable Register of the status model.
Type Affected by Where saved Default value
Command and query
STATus:PRESet
Nonvolatile memory
0
Usage
:STATus:OPERation:ENABle <n>
:STATus:OPERation:ENABle?
<n>
The status of the operation status register

Table of Contents

Other manuals for Keithley 2450

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Keithley 2450 and is the answer not in the manual?

Keithley 2450 Specifications

General IconGeneral
BrandKeithley
Model2450
CategoryTest Equipment
LanguageEnglish

Related product manuals