6: SCPI command reference Model 2450 Interactive SourceMeter® Instrument
6-120 2450-901-01 Rev. B/September 2013
Details
This command returns the average reading calculated from all of the readings in the specified reading buffer.
Example
Returns the average reading for the readings in the default
buffer
.
TRACe:STAT:AVERage? "testData"
Returns the average reading for the readings in the user-
defined buffer
.
Also see
Reading buffers (on page 3-10)
Remote buffer operation (on page 3-27)
:TRACe:MAKE (on page 6-114)
:TRACe:STATistics:CLEar (on page 6-120)
:TRACe:STATistics:MAXimum? (on page 6-121)
:TRACe:STATistics:MINimum? (on page 6-121)
:TRACe:STATistics:PK2Pk? (on page 6-122)
:TRACe:STATistics:STDDev? (on page 6-123)
:TRACe:STATistics:CLEar
This command clears the statistical information associated with the specified buffer.
Type Affected by Where saved Default value
Usage
TRACe:STATistics:CLEar
TRACe:STATistics:CLEar <bufferName>
The name of the reading buffer, which may be a default buffer
(defbuffer1 or defbuffer2) or a user-defined buffer
Example
Clear all statistics in defbuffer1.
TRACe:STATistics:CLEar "testData"
Clears all statistics in a user-defined buffer
named testData.
Also see
Reading buffers (on page 3-10)
Remote buffer operation (on page 3-27)
:TRACe:MAKE (on page 6-114)
:TRACe:STATistics:AVERage? (on page 6-119)
:TRACe:STATistics:MAXimum? (on page 6-121)
:TRACe:STATistics:MINimum? (on page 6-121)
:TRACe:STATistics:PK2Pk? (on page 6-122)
:TRACe:STATistics:STDDev? (on page 6-123)