Interactive SourceMeter® Instrument Reference Manual Section 6: SCPI command refe
2450-901-01 Rev. B/September 2013 6-147
:TRIGger:BLOCk:SOURce:STATe
This command defines a trigger block that turns the output source on or off.
Type Affected by Where saved Default value
Instrument reset
Usage
:TRIGger:BLOCk:SOURce:STATe <blockNumber>, <state>
The sequence of the block in the trigger model
Disable the source:
Details
The source building block determines if the output source is turned on or off when the trigger model
reaches this block.
This block does not determine the settings of the output source (such as the output voltage level and
source delay). The source settings are determined by either the present settings of the instrument or
by a source configuration list.
When you list trigger blocks, this block is listed as SOURCE_OUTPUT.
Example
TRIG:BLOC:DEL:CONS 2, 0.01
TRIG:BLOC:MEAS 3
TRIG:BLOC:BRAN:COUN 4, 100, 2
TRIG:BLOC:SOUR:STAT 5, 0
This example turns the output on.
Delay 10 ms.
Make a measurement.
Loop and take 100 readings.
The output is turned off after the 100 readings
Also see
Wait building block (on page 3-96)