Interactive SourceMeter® Instrument Reference Manual Section 8:
2450-901-01 Rev. B/September 2013 8-233
tsplink.line[N].reset()
This function resets some of the TSP-Link trigger attributes to their defaults.
Type TSP-Link accessible
Affected by Where saved Default value
Usage
tsplink.line[N].reset()
The trigger line (1 to 3)
Details
The tsplink.line[N].reset() function resets the following attributes to their default values:
• trigger.tsplinkin[N].edge
• trigger.tsplinkout[N].logic
• tsplink.line[N].mode
• trigger.tsplinkout[N].stimulus
• trigger.tsplinkout[N].pulsewidth
This also clears trigger.tsplinkin[N].overrun.
Example
Resets TSP-Link trigger line 3 attributes to default values.
Also see
trigger.tsplinkin[N].edge (on page 8-225)
trigger.tsplinkin[N].overrun (on page 8-226)
trigger.tsplinkout[N].logic (on page 8-228)
trigger.tsplinkout[N].pulsewidth (on page 8-228)
trigger.tsplinkout[N].stimulus (on page 8-229)
tsplink.line[N].mode (on page 8-232)
tsplink.line[N].state
This attribute reads or writes the digital state of a TSP-Link synchronization line.
Type TSP-Link accessible Affected by Where saved Default value
Usage
lineState = tsplink.line[N].state
tsplink.line[N].state = lineState
The state of the synchronization line:
• Low: tsplink.STATE_LOW or 0
•
High: tsplink.STATE_HIGH or 1
The trigger line (1 to 3)