Section
8: TSP command reference Model 2450 Interactive SourceMeter® Instrument
Reference Manual
8-110 2450-901-01 Rev. B/September 2013
smu.measure.limit[Y].low.value
This command specifies the lower limit for limit tests.
Type TSP-Link accessible Affected by Where saved Default value
Attribute (RW) Yes
Restore configuration
Instrument reset
Power cycle
Measure configuration list
Function change
Configuration script
Measure configuration list
1.000000E+00
Usage
value = smu.measure.limit[Y].low.value
smu.measure.limit[Y].low.value = value
value
The high value; range is 999999900000 to
+
999999900000
Y
Limit number: 1 or 2
Details
This command sets the lower limits for the limit tests for the selected measurement function. When limit Y testing
is enabled, this causes a fail indication to occur when the measurement value is less than this value.
Example
See the example in smu.measure.limit[Y].fail (on page 8-107).
Also see
smu.measure.limit[Y].enable (on page 8-106)

Table of Contents

Other manuals for Keithley 2450

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Keithley 2450 and is the answer not in the manual?

Keithley 2450 Specifications

General IconGeneral
BrandKeithley
Model2450
CategoryTest Equipment
LanguageEnglish

Related product manuals