8: TSP command reference Model 2450 Interactive SourceMeter® Instrument
8-110 2450-901-01 Rev. B/September 2013
smu.measure.limit[Y].low.value
This command specifies the lower limit for limit tests.
Type TSP-Link accessible Affected by Where saved Default value
Attribute (RW) Yes
Restore configuration
Instrument reset
Power cycle
Measure configuration list
Configuration script
Measure configuration list
−1.000000E+00
Usage
value = smu.measure.limit[Y].low.value
smu.measure.limit[Y].low.value = value
The high value; range is −999999900000 to
999999900000
Details
This command sets the lower limits for the limit tests for the selected measurement function. When limit Y testing
is enabled, this causes a fail indication to occur when the measurement value is less than this value.
Example
See the example in smu.measure.limit[Y].fail (on page 8-107).
Also see
smu.measure.limit[Y].enable (on page 8-106)