8: TSP command reference Model 2450 Interactive SourceMeter® Instrument
Reference Manual
8-142 2450-901-01 Rev. B/September 2013
smu.source.protect.tripped
This attribute indicates if the overvoltage source protection feature is active.
Type TSP-Link accessible Affected by Where saved Default value
Attribute (R) Yes Not applicable Not applicable Not applicable
Usage
value = smu.source.protect.tripped
value
Overvoltage protection not activated:
smu.OFF
Overvoltage protection activated: smu.ON
Details
When overvoltage protection is active, the instrument restricts the maximum voltage level that the instrument can
source.
Example
print(smu.source.protect.tripped)
If overvoltage protection is active, the output
is:
smu.ON
Also see
smu.source.protect.level (on page 8-141)