Section
8: TSP command reference Model 2450 Interactive SourceMeter® Instrument
Reference Manual
8-168 2450-901-01 Rev. B/September 2013
Details
The trigger.blender[N].reset() function resets the following attributes to their factory defaults:
trigger.blender[N].orenable
trigger.blender[N].stimulus[M]
It also clears trigger.blender[N].overrun.
Example
trigger.blender[1].reset()
Resets the trigger blender 1
settings to factory defaults.
Also see
trigger.blender[N].orenable (on page 8-166)
trigger.blender[N].overrun (on page 8-167)
trigger.blender[N].stimulus[M] (on page 8-168)
trigger.blender[N].stimulus[M]
This attribute specifies which events trigger the blender.
Type TSP-Link accessible Affected by Where saved Default value
Attribute (RW)
Yes
Restore configuration
Instrument reset
Power cycle
Trigger blender N reset
Configuration script
trigger.EVENT_NONE
Usage
event = trigger.blender[N].stimulus[M]
trigger.blender[N].stimulus[M] = event
event
The event that triggers the blender action; see Details
N
An integer that represents the trigger event blender (1 or 2)
M
An integer representing the stimulus index (1 to 4)
Details
There are four stimulus inputs that can each select a different event. The event parameter can be any trigger
event.
Use zero to disable the blender input.

Table of Contents

Other manuals for Keithley 2450

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Keithley 2450 and is the answer not in the manual?

Keithley 2450 Specifications

General IconGeneral
BrandKeithley
Model2450
CategoryTest Equipment
LanguageEnglish

Related product manuals