8: TSP command reference Model 2450 Interactive SourceMeter® Instrument
Reference Manual
8-106 2450-901-01 Rev. B/September 2013
smu.measure.limit[Y].enable
This attribute enables or disables a limit test.
Type TSP-Link accessible Affected by Where saved Default value
Attribute (RW) Yes
Restore configuration
Instrument reset
Power cycle
Measure configuration list
Function change
Configuration script
Measure configuration list
smu.OFF
Usage
state = smu.measure.limit[Y].enable
smu.measure.limit[Y].enable = state
state
Disable the test:
smu.OFF
or
0
Enable the test: smu.ON or 1
Y
The number of the test: 1 or 2
Details
This command enables or disables a limit test for the selected measurement function.
Example
smu.measure.func = smu.FUNC_DC_VOLTAGE
smu.measure.limit[1].enable = smu.ON
Enable testing for limit 1 when measuring
voltage.
Also see
smu.measure.limit[Y].autoclear (on page 8-104)
smu.measure.limit[Y].clear() (on page 8-105)
smu.measure.limit[Y].fail (on page 8-107)
smu.measure.limit[Y].high.value (on page 8-109)
smu.measure.limit[Y].low.value (on page 8-110)

Table of Contents

Other manuals for Keithley 2450

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Keithley 2450 and is the answer not in the manual?

Keithley 2450 Specifications

General IconGeneral
BrandKeithley
Model2450
CategoryTest Equipment
LanguageEnglish

Related product manuals