Interactive SourceMeter® Instrument Reference Manual Section 6:
SCPI command reference
2450-901-01 Rev. B/September 2013 6-61
:SOURce[1]:CONFiguration:LIST:CREate
This command creates an empty source configuration list for sourcing.
Type Affected by Where saved Default value
Command only
Recall settings
Instrument reset
Power cycle
Save settings
Not applicable
Usage
:SOURce[1]:CONFiguration:LIST:CREate <name>
<name>
A string that represents the name of a source configuration list
Details
This command creates an empty configuration list. To add configuration points to this list, you need to use the
store command.
Configuration lists are not saved when the instrument is turned off. If you want to save a configuration list, use a
saved setup to store the instrument settings, which include defined configuration lists.
Examples
:SOURce:CONF:LIST:CRE "MySourceList"
Creates a source configuration list named MySourceList.
Also see
Configuration lists (on page 3-33)
*SAV (on page 6-2)
:SOURce[1]:CONFiguration:LIST:STORe (on page 6-65)

Table of Contents

Other manuals for Keithley 2450

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Keithley 2450 and is the answer not in the manual?

Keithley 2450 Specifications

General IconGeneral
BrandKeithley
Model2450
CategoryTest Equipment
LanguageEnglish

Related product manuals