Model 2450
Interactive SourceMeter® Instrument Reference Manual Section 6:
SCPI command reference
2450-901-01 Rev. B/September 2013 6-39
[:SENSe[1]]:CONFiguration:LIST:QUERy?
This command returns a list of TSP commands that represent the parameters that are stored in the specified
configuration point.
Type Affected by Where saved Default value
Query only
Not applicable
Not applicable
Not applicable
Usage
[:SENSe[1]]:CONFiguration:LIST:QUERy? <name>, <point>
<name>
A string that represents the name of a measure configuration list
<point>
A specific configuration point in the configuration list
Details
This command can only return data for one configuration point. To get data for additional configuration points,
send the command specifying different configuration points each time.
For additional information about the attributes this command returns, see Instrument settings stored in a
measure configuration list (on page 3-36).
Example
:SENS:CONF:LIST:QUER? "MyMeasList", 2
Returns the TSP commands that represent the settings in configuration point 2.
Also see
Configuration lists (on page 3-33)
[:SENSe[1]]:CONFiguration:LIST:CREate (on page 6-37)

Table of Contents