Model 2450
Interactive SourceMeter® Instrument Reference Manual Section 8:
TSP command reference
2450-901-01 Rev. B/September 2013 8-151
smu.source.sweeplog()
This function creates a logarithmic source sweep configuration list and trigger model.
Type TSP-Link accessible
Affected by Where saved Default value
Function
Yes
Usage
smu.source.sweeplog(configListName, start, stop, points)
smu.source.sweeplog(configListName, start, stop, points, delay)
smu.source.sweeplog(configListName, start, stop, points, delay, count)
smu.source.sweeplog(configListName, start, stop, points, delay, count, rangeType)
smu.source.sweeplog(configListName, start, stop, points, delay, count, rangeType,
failAbort)
smu.source.sweeplog(configListName, start, stop, points, delay, count, rangeType,
failAbort, dual)
smu.source.sweeplog(configListName, start, stop, points, delay, count, rangeType,
failAbort, dual, bufferName)
smu.source.sweeplog(configListName, start, stop, points, delay, count, rangeType,
failAbort, dual, bufferName, asymptote)
configListName
A string that contains the name of the configuration list that the instrument will
create for this sweep
start
The voltage or current source level at which the sweep starts:
Current: 1 pA to 1.05
Voltage: 1 pV to 210
stop
The voltage or current at which the sweep stops:
Current: 1 pA to 1.05
Voltage: 1 pV to 210
points
The number of source-measure points between the start and stop values of the
sweep (2 to 1e6); to calculate the number of source-
measure points in a sweep, use
the following formula:
Points = [(Stop - Start) / Step] + 1
delay
The delay between measurement points; default is smu.DELAY_AUTO, which
enables autodelay, or a specific delay value from 50 µs to 10,000 seconds, or 0 for
no delay
count
The number of times to run the sweep; default is 1:
Infinite loop: smu.INFINITE
Finite loop:
1
to
268435455
rangeType
The source range that is used for the sweep:
Most sensitive source range for each source level in the sweep:
smu.RANGE_AUTO or 0
Best fixed range: smu.RANGE_BEST or 2 (default)
Present source range for the entire sweep: smu.RANGE_FIXED or 1
failAbort
Abort the sweep if the source limit is exceeded: smu.ON (default)
Complete the sweep if the source limit is exceeded: smu.OFF
dual
Determines if the sweep runs from start to stop and then from stop to start:
Sweep from start to stop only: smu.OFF (default)
Sweep from start to stop, then stop to start: smu.ON
bufferName
The name of a reading buffer; the default buffers (defbuffer1 or defbuffer2) or
the name of a user-
defined buffer; if no buffer is specified, this parameter defaults to
defbuffer1

Table of Contents

Other manuals for Keithley 2450

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Keithley 2450 and is the answer not in the manual?

Keithley 2450 Specifications

General IconGeneral
BrandKeithley
Model2450
CategoryTest Equipment
LanguageEnglish

Related product manuals