EasyManuals Logo
Home>Keithley>Test Equipment>2450

Keithley 2450 User Manual

Keithley 2450
878 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Page #442 background imageLoading...
Page #442 background image
Model 2450
Interactive SourceMeter® Instrument Reference Manual Section 6:
SCPI command reference
2450-901-01 Rev. B/September 2013 6-125
:TRIGger:BLENder<n>:CLEar
This command clears the blender event detector and resets the overrun indicator of blender <n>.
Type Affected by Where saved Default value
Command only Not applicable Not applicable Not applicable
Usage
:TRIGger:BLENder<n>:CLEar
<n>
The blender number (1 or 2)
Details
This command sets the blender event detector to the undetected state and resets the overrun indicator of the
event detector.
Example
:TRIG:BLEN2:CLE
Clears the event detector for blender 2.
Also see
None
:TRIGger:BLENder<n>:MODE
This command selects whether the blender performs OR operations or AND operations.
Type Affected by Where saved Default value
Command and query
Recall settings
Instrument reset
Power cycle
Trigger blender clear
Save settings
AND
Usage
:TRIGger:BLENder<n>:MODE <operation>
:TRIGger:BLENder<n>:MODE?
<n>
The blender number (1 or 2)
<operation>
The type of operation:
OR
AND

Table of Contents

Other manuals for Keithley 2450

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Keithley 2450 and is the answer not in the manual?

Keithley 2450 Specifications

General IconGeneral
BrandKeithley
Model2450
CategoryTest Equipment
LanguageEnglish

Related product manuals