8: TSP command reference Model 2450 Interactive SourceMeter® Instrument
8-46 2450-901-01 Rev. B/September 2013
Example
print(data)
Assume lines 2, 4, and 6 are set high when the
I/O port is read.
Output:
42
Also see
digio.writeport() (on page 8-46)
Digital I/O port (on page 3-62)
digio.writeport()
This function writes to all digital I/O lines.
Type TSP-Link accessible
Affected by Where saved Default value
Usage
digio.writeport(data)
The value to write to the port (0 to 63)
Details
The binary representation of the value indicates the output pattern to be written to the I/O port. For example, a
value of 63 has a binary equivalent of 111111 (all lines are set high); a data value of 42 has a binary equivalent
of 101010 (lines 2, 4, and 6 are set high, and the other 3 lines are set low).
An instrument reset does not affect the present states of the digital I/O lines.
Example
Sets digital I/O lines 1 through 6 high (binary
Also see
digio.readport() (on page 8-45)
Digital I/O port (on page 3-62)