8: TSP command reference Model 2450 Interactive SourceMeter® Instrument
8-184 2450-901-01 Rev. B/September 2013
Details
The LAN trigger listens for trigger messages on all the supported protocols. However, it uses the designated
protocol for sending outgoing messages.
After you change this setting, you must re-connect the LAN trigger event generator before you can send outgoing
event messages.
When multicast is selected, the trigger IP address is ignored and event messages are sent to the multicast
address 224.0.23.159.
Example
print(trigger.lanout[1].protocol)
Get LAN protocol that is being used for sending
trigger messages for LAN event 1.
Also see
trigger.lanout[N].connect() (on page 8-180)
trigger.lanout[N].ipaddress (on page 8-182)
trigger.lanout[N].stimulus
This attribute specifies events that cause this trigger to assert.
Type TSP-Link accessible Affected by Where saved Default value
Instrument reset
Power cycle
Usage
event = trigger.lanout[N].stimulus
trigger.lanout[N].stimulus = event
The LAN event that causes this trigger to assert
A number specifying the trigger packet over the LAN for which to set or query
the trigger source (1 to 8)