8: TSP command reference Model 2450 Interactive SourceMeter® Instrument
Reference Manual
8-208 2450-901-01 Rev. B/September 2013
trigger.model.setblock() trigger.BLOCK_LOG_EVENT
This function allows you to log an event in the event log when the trigger model is running.
Type TSP-Link accessible Affected by Where saved Default value
Function
Yes
Restore configuration
Instrument reset
Power cycle
Configuration script
Not applicable
Usage
trigger.model.setblock(blockNumber, trigger.BLOCK_LOG_EVENT, eventNumber, message)
blockNumber
The sequence of the block in the trigger model
eventNumber
The event number:
trigger.LOG_INFON
trigger.LOG_WARNN
trigger.LOG_ERRORN
Where
N
is 1 to 4; you can define up to four of each type
message
A string up to 31 characters
Details
This block allows you to log an event in the event log when the trigger model is running. Insert the block into the
trigger model. When the trigger model executes the block, the event is logged.
Note that using this block too often in a trigger model could overflow the event log. It may also take away from
the time needed to process more critical trigger model blocks.
Example
trigger.model.setblock(1, trigger.BLOCK_LOG_EVENT, trigger.LOG_INFO1, "info
message")
Set trigger model block 1 to log an event.
Also see
None

Table of Contents

Other manuals for Keithley 2450

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Keithley 2450 and is the answer not in the manual?

Keithley 2450 Specifications

General IconGeneral
BrandKeithley
Model2450
CategoryTest Equipment
LanguageEnglish

Related product manuals