3: Functions and features Model 2450 Interactive SourceMeter® Instrument
3-38 2450-901-01 Rev. B/September 2013
Instrument settings stored in a source configuration list
When you save a configuration point to a source configuration list, the instrument saves the values for
each setting listed in the following table. These settings can be set from the front panel or by using
remote commands. The table lists SCPI and TSP commands that can be used to set these settings.
Source configuration list settings
Front-panel setting TSP command SCPI command
smu.source.func (on page 8-
:SOURce[1]:FUNCtion[:MODE] (on page 6-72)
AutoRange
HOME > Source Range
smu.source.autorange (on
page 8-127)
:SOURce[1]:<function>:RANGe:AUTO (on page 6-75)
MENU > Source > Settings > Source
smu.source.delay (on page 8-
133)
:SOURce[1]:<function>:DELay (on page 6-66)
Auto delay
Not available from the front panel
smu.source.autodelay (on
page 8-128)
:SOURce[1]:<function>:DELay:AUTO (on page 6-67)
HOME > Source
smu.source.level (on page 8-
135)
:SOURce[1]:<function>:<x>LIMit[:LEVel] (on page 6-71)
MENU > Source > Settings >
Overvoltage Protection Limit
smu.source.protect.level (on
page 8-141)
:SOURce[1]:<function>:PROTection[:LEVel] (on page 6-
73)
MENU > Source > Settings > Output
smu.source.offmode (on page
8-139)
:OUTPut[1]:<function>:SMODe (on page 6-31)
MENU > Source > Settings > Source
smu.source.range (on page
8-143)
:SOURce[1]:<function>:RANGe (on page 6-74)
HOME > Limit
smu.source.xlimit.level (on
page 8-137)
:SOURce[1]:<function>:<x>LIMit[:LEVel] (on page 6-71)
MENU > Source > Settings > Hi
smu.source.highc (on page 8-
135)
:SOURce[1]:<function>:HIGH:CAPacitance (on page 6-
69)
MENU > Source > Settings > Source
Readback
page 8-144)
:SOURce[1]:<function>:READ:BACK (on page 6-77)
User delays
Not available from the front panel
smu.source.userdelay[N] (on
page 8-153)
:SOURce[1]:<function>:DELay:USER<n> (on page 6-68)
Source output
Output on/off switch
smu.source.output (on page
8-140)
:OUTPut[1][:STATe] (on page 6-34)