6: SCPI command reference Model 2450 Interactive SourceMeter® Instrument
6-44 2450-901-01 Rev. B/September 2013
[:SENSe[1]]:<function>:AVERage[:STATe]
This command enables or disables the averaging filter for measurements of the selected function.
Type Affected by Where saved Default value
Command and query
Recall settings
Instrument reset
Power cycle
Measure configuration list
Save settings
Measure configuration list
0 (OFF)
Usage
[:SENSe[1]]:<function>:AVERage[:STATe] <state>
[:SENSe[1]]:<function>:AVERage[:STATe]?
The measurement function to which this setting applies:
• Current: CURRent[:DC]
• Resistance: RESistance
• Voltage:
The filter status; set to one of the following values:
• Disable the averaging filter: 0 or OFF
• Enable the averaging filter:
or
Details
This command enables or disables the averaging filter. When this is enabled, the measurements for the selected
measurement function are averaged as set by the filter count and filter type.
If you send this command without the <function> parameter, it sets the state of the averaging filter
for all functions.
Example 1
CURR:AVER:TCON MOV
Set the averaging filter type to moving
average, with a filter count of 10.
Enable the averaging filter.
Example 2
RES:AVER:TCON MOV
Set the averaging filter type to moving
average, with a filter count of 10.
Enable the averaging filter.
Example 3
VOLT:AVER:TCON MOV
Set the averaging filter type to moving
average, with a filter count of 10.
Enable the averaging filter.
Also see
Filtering measurement data (on page 4-22)
[:SENSe[1]]:<function>:AVERage:COUNt (on page 6-43)
[:SENSe[1]]:<function>:AVERage:TCONtrol (on page 6-45)