93
3.1 Maintenance Items
REGIUS MODEL 190 Service Manual Ver.1.00 2004.11.01
3.1 Maintenance Items
3.1.1
Maintenance & Service Schedule
Carry out maintenance and service following the timetable shown below.
Maintenance by user: Refer to the operation manual for details.
Other maintenance by qualied person only
Frequency
Maintenance Task
Weekly Monthly Every 6 mo.
Cleaning of the cassette stacker
Cleaning of the REGIUS Plate
Cleaning of the outer covers
Cleaning of the exhaust slot
Frequency
Maintenance Task
Every 6 mo. Every 12 mo. Refer to;
Cleaning of the cassette magnetizing
(absorption) plate
p.98
Cleaning of the exit rollers
Lubrication to the LM guide
p.94
Greasing the warm wheel of transport
unit
p.95
Lubrication to the release shaft holder
p.96
Cleaning of the power supply cooling
fan
p.97
Replacement of the erasing lamps
app. every 10,000 shots p.78